MQ-6000

MQ-6000

Principle

  • High Performance Liquid Chromatography (HPLC) 

Parameters

  • Global Standard: MQ-6000 Glycated hemoglobin analyzer uses gold-standard, high performance liquid chromatography (HPLC) to deliver precise results in Hb1Ac

  • IFCC Traceable Results: Calibrators value traceable to IFCC-RMP, Lyophylized calibration and quality control materials allow for long shelf life and easy transportation

  • Technology Assurance: Self-developed high pressure plunger pump (Micro-pump) ensures constant liquid velocity, giving accurate results. 24 bit AD high precision detector can detect sample absorbance with high resolution

Please Log In to access Additional Details and documents ..

Fully-automatic Features

  • Automatic identification of sample type Whole blood: 10uL, Loading 100 samples at time

  • Automatic barcode reading and puncture

  • Automatic mixing, prevent sample layering

  • STAT Position: 1 STAT Position allows high priority sample to be inserted any time

  • One-Reagent - Two modes: No additional reagents are required to switch variant test mode. Hb-Variant detection: HbD, HbS, HbC, HbE

Interference avoidance 

  • No clinical interference of HbD, HbE, HbC and HbS

  • No clinical interference with labite A1c, acetylated Hb, aldehyde Hb and carbomylated Hb

Specification 

  • Analytes: HbA1c, HbE, HbF, HbD, HbS,  HbA1a, LA1c, HbA0

  • Throughput: 72 sec/test for normal mode, 110 sec/test for variant mode 

  • Sample Consumption: Whole Blood 10uL

  • Loading Capacity: 100 Samples

  • Precision: CV <= 2%

  • Measuring Range (NGSP): 4.0% - 17.0%

  • Data Storage: On-board memory up to 100,000 results with graphs

  • Display: 10’1 inch touch screen

  • Output: Internal Thermal Printer, LIS, USB

  • Sample Holder: Dilution and Whole Blood Sample Auto-switching, Internal barcode reader, Auto-mixing, Cap Piercing

  • Working Environment: 10C - 30C Humidity <= 70% Altitude: self-adaptive

  • Dimensions: 540 mm x 540 mm x 540 mm (L*W*H)

  • Weight 40.3 kg